GC-7800 with autosampler is a classical and high automation gas chromatograph instrument which integrates the most advanced chromatography technique and national or international standards with the most extensive applications by the experienced experts of Labthink. Its automatic preparing, sampling and analyzing functions effectively avoid system errors caused by manual operations. This instrument can be widely used to test the relative indexes of various materials, gases, smell, residue, and cigarette packages, because of its powerful functions and ultra high precision. It becomes the best choice for gas chromatographic analysis and control of testing organizations, research institutions, packaging companies as well as food and pharmaceutical industries.
Technical Specifications
Temperature of Specimen Chamber A | Room temperature +5°C ~ 200°C with ±0.1°C accuracy |
---|
Temperature of Specimen Chamber B | Room temperature +5°C ~ 200°C with ±0.1°C accuracy |
---|
Temperature of Specimen Gas Path | Room temperature +5°C ~ 150°C with ±0.5°C accuracy |
---|
Temperature of Auto-Feeding Specimen Switch Box | Room temperature +5°C ~ 150°C with ±0.5°C accuracy |
---|
Specimen Volume | 1.0 mL (0.5 mL, 2 mL, or 3 mL can be customized) |
---|
Headspace Bottle Specification | 300 mL (Standard) |
---|
Specimen Analysis Station | 2 |
---|
Counter Blowing Cleaning | 0 ~ 60000 Seconds |
---|
Specimen-Feeding Time | 0 ~ 60000 Seconds |
---|
Headspace Balance Heating Time | 0 ~ 60000 Seconds |
---|
Specimen-Feeding Switch Time | 0 ~ 60000 Seconds |
---|
Temperature of Chromatographic Column Chamber | Temperature Accuracy Better than ±0.1°C |
---|
Temperature Gradient Effective range of column is less than 1% |
Temperature Deviation Deviation between setting temperature and display temperature is less than 1°C |
Temperature Deviation Deviation between setting temperature and actual temperature is less than 2% |
Orders of Temperature Program | 5 orders (customization available) |
---|
Rate of Temperature | 0.1 ~ 30°C/min |
---|
Temperature Range of Linear Temperature Program | Lower than 150 °C while 30°C per minute Lower than 300 °C while 15°C per minute Lower than 350 °C while 10°C per minute |
---|
Control Time from Primary to Terminal Temperature | 0 ~ 600 min |
---|
Repeatability of Temperature Program | Less than 2% |
---|
Temperature Decreasing Speed | Less than 15min from 300°C to 50°C (While temperature of vaporizing and test chambers increases to 300 °C) |
---|
Vaporizing Chamber | Temperature Accuracy ±0.1°C (room temperature +15°C ~ 200°C) ±0.2°C while temperature is higher than 200°C |
---|
Test Chamber | Temperature Accuracy ±0.1°C (room temperature +15°C ~ 200°C) ±0.2°C while temperature is higher than 200°C |
---|
Flame Ionization Detector (FID) | Test Limit Less than 1 x 10-11g/s (Benzene) |
---|
Noise Less than 0.025 mV |
Drift Less than 0.15 mV/h |
Thermal Conductivity Detector (TCD) (Optional Parts) | Sensitivity More than 3000 mV·mL/mg (Benzene) |
---|
Noise Less than 0.035 mV |
Drift Less than 0.5 mV/h |
Mainframe | Power Supply AC 220V 50Hz |
---|
Instrument Dimension 620 mm (L) x 490 mm (W) x 480 mm (H) |
Net Weight 55 kg |
Auto Sampler | Instrument Dimension 340 mm (L) x 535 mm (W) x 495 mm (H) |
---|
Net Weight 27 kg |
A gas chromatograph is a separating or analytical instrument for mixture containing various components. It takes the inert gas as mobile phase, and utilizes the chromatographic column separation technology. The mixture injected into the instrument follows the carrier gas into a separation tube known as "column". Because of different distribution coefficients in the column, distinct components will display different mobile speeds. After certain column, the analyzed components will leave the column consecutively into the detector where corresponding electrical signals will be generated and taken into chromatography data system. The quantitative analysis of tested mixture is accomplished.
Professional Technology
Dual-detector (FID+TCD), dual-column and dual-gas path systemInstrument utilizes patent design of solvent residue test, and is available for the determination of solvent purity and water content of hydrophilic solvents.This instrument supports the organic compound determination of packing paper for cigarette boxes and label paper.Utilizing patent design of automatic preparing and feeding sampleControlled by micro-computerTest parameters are displayed by LCDEquipped with two constant temperature chambers for sample preparationTwo specimens can be tested simultaneously Special design of anti-condensation and back-flush cleaning avoids cross contamination between specimensVarious testing modes of constant temperature, temperature program and mixed control are availableLinear temperature program with 5 orders (customization available)Power off protection against over- or drop- temperatureModular designed detector provides extendable functionsPractical design of base line auto-zero, power failure protection, and test data storage or inquiry ensures the safe operation conditionSupported by professional chromatography data system with powerful processing functions Characteristics of Gas Chromatograph
The instrument is controlled by micro-computer and integrated circuit, with big display screen. Functions of keyboard inputting parameters, power failure protection and file storage or inquiry provide a comfortable operating environment.Reliable and flexible gas path system is easy to be extended with dual column and dual gas path operation, which is available for different feeding gases or tests and selected in accordance with user analysis requirements.Detectors uses unitized combination design and plug-and-play (PNP) extension control mode. Instrument is equipped with hydrogen flame ionization detector (FID) and thermal conductivity detector (TCD). Four different detectors could be installed simultaneously at most in accordance with user analysis requirements.Constant temperature or temperature program can be achieved in this instrument. And preset temperature of chromatographic column chamber could be reached through automatic open-close of its flexible back door.Sampler system is equipped with packed column, capillary column and other devices to transfer test gases into the system.Instrument is designed with functions of lower pressure for carrier gas tube and protection for gas transmission interrupt.High temperature protection is applied so that instrument could stop working and alarm when the temperature of any gas tube exceeds the preset temperature (set at 20°C).1500mv wide range of response signal, with good consistency, could meet analysis requirements of high purity samples.Good performance of chromatographic column box makes it possible to achieve the operation at room temperature+3°C, even when the temperatures of detection and vaporizing chambers are as high as 200°C. Instrument comes with good repeatability of temperature program to ensure the retention time consistency of sample components.Special designed operating software and hardware could automatically monitor the whole operating process, which creates a safe test environment for users.Instrument could be easily connected with chromatography data converter, data system or other graphing devices. Professional chromatographic data system provides powerful and intelligent data process function to automatically identify the names and quantities (mg/m2) of residual solvents of tested samples in accordance with reference data. Instrument is also available for the detection of solvent purity and water content of hydrophilic solvents.Characteristics of Auto Sampler
Controlled by single chip microcomputer, with keyboard for parameter input, LCD display; high automation provides convenient operating experience.Auto sampler utilizes integrated structure design, with 300mL sample bottle. Standard specimen chamber is totally separated from test specimen chamber, so that parameter setting and heating could be individually performed.Feeding sample valve and transmission tube can be heated to preset temperature to avoid sample condensation.Back-flush cleaning of feeding valve and tube avoids cross contamination between specimensCombination of sampling pump, feeding valve and negative pressure method provides constant test pressure, accurate quantify, good repeatability and high sensitivity.Equipped with communication port which is used to connect gas chromatograph and chromatography data system to accomplish the one-click operation of preparing, sampling, and analyzing.